FirsthandHealth
FDA MedWatch Safety AlertsUnited States19 November 2026

React Health Recalls 7,533 Luna G3 BPAP Devices Over Pressure Data Error

On November 19, 2026, the US Food and Drug Administration published a safety notice regarding a voluntary medical device correction by React Health (3B Medical, Inc.). The company recalled 7,533 Luna G3 Bilevel Positive Airway Pressure (BPAP) System devices, specifically Model LG3800-30VT, distributed across the US and Puerto Rico. When operated with the Targeted VT feature enabled, these devices fail to record or transmit actual therapy pressure statistical data to internal SD cards or the cloud platform. React Health has received 15 complaints and two reports of serious adverse events connected to this issue. Inaccurate historical pressure reports could cause doctors to make unnecessary or incorrect adjustments to treatment settings, potentially delaying proper care. Patients should not stop their therapy or change settings independently. Instead, they should contact their healthcare provider or equipment supplier to arrange a free firmware update, version G3-2.00.80.

What was announced

React Health initiated a voluntary recall for specific units of its Luna G3 BPAP System, Model LG3800-30VT. The US Food and Drug Administration published the notification on November 19, 2026. The correction affects 7,533 devices distributed across the US and Puerto Rico.

The issue occurs when the machine operates with the Targeted VT feature enabled. Under these conditions, the device fails to save actual therapy pressure data onto its internal SD card. It also fails to transmit pressure data to React Health Connect, the cloud management system. Devices operating without Targeted VT enabled do not experience this tracking error.

React Health created firmware version G3-2.00.80 to correct the reporting issue. The company is providing update SD cards through medical equipment providers so devices can be updated without full replacement.

What it means

While the device continues to deliver air within the set minimum and maximum pressure limits, the inaccurate historical data can mislead medical staff. Clinicians reviewing therapy logs will see non-fluctuating pressure readings that do not reflect what the patient received.

This faulty reporting can lead clinicians to alter treatment settings incorrectly, which may delay proper respiratory therapy until corrected. React Health reported receiving 15 complaints and two reports of serious adverse events tied to this issue.

Patients using an affected device should continue their prescribed treatment. They must not turn off the machine or adjust settings without medical advice. Patients should contact their doctor or equipment provider to install the firmware update. Clinicians must immediately stop using Targeted VT on affected models and evaluate patient therapy histories carefully.

What the announcement does not tell us

The announcement does not describe the specific injuries or health impacts reported in the two serious adverse event filings. It also does not state when React Health manufactured or sold the affected machines.

Additionally, the document does not specify how long it will take React Health to ship firmware SD cards to all equipment suppliers and medical facilities. It leaves unclear whether patients who received delayed therapy will require additional medical testing or specialized follow-up assessments.

How this was written

This analysis was written with AI from the official announcement below, using only facts that announcement states. The quoted wording is FDA MedWatch Safety Alerts’s own. Always check the original before acting on anything here.

React Health (3B Medical, Inc.) is providing notice of a voluntary medical device correction (recall) involving specific Luna G3 BPAP System (Model LG3800-30VT) devices due to a data handling issue that may result in incorrect therapy pressure statistical data when the device is operated with Target
Source: FDA MedWatch Safety Alerts

More from FDA MedWatch Safety Alerts

This content is for informational purposes only and is not medical advice. It is not intended to diagnose, treat, cure, or prevent any disease. Consult a healthcare professional before starting any supplement, treatment, or program: especially if you are pregnant, nursing, taking medication, or managing a health condition.