FDA Issues Safety Alert for Outset Medical Tablo Dialysis Power Cords Overheating
The U.S. Food and Drug Administration issued an early alert on November 19, 2026, regarding overheating risks with Outset Medical Tablo Hemodialysis Systems. Outset Medical sent a safety letter to customers on November 10, 2026, after receiving reports of heat generating where the power cord plugs into the Tablo Console and TabloCart with Prefiltration Drawer. Hemodialysis is a medical treatment that filters waste and excess fluid from the blood when the kidneys fail. The issue affects power cords with both clear ends (part number PN-0005746) and gray ends (part number PN-0008460). A loose power cord connection can increase electrical resistance, causing extreme heat, smoke, sparks, or loss of power. If the connection overheats, a person touching the connector could suffer first- or second-degree burns. Additionally, power loss could interrupt or delay kidney dialysis treatments. As of November 10, 2026, Outset Medical reported zero serious injuries or deaths linked to the issue.
What was announced
On November 19, 2026, the Center for Devices and Radiological Health at the U.S. Food and Drug Administration posted an early alert about Outset Medical Tablo Hemodialysis Systems. The warning covers the Tablo Console and the TabloCart with Prefiltration Drawer. Specifically, the alert identifies two power cords: part number PN-0005746 with clear ends and part number PN-0008460 with gray ends.
On November 10, 2026, Outset Medical notified customers to follow updated instructions. The company received reports of overheating where the power cord meets the power inlet. When the power cord connector is not held firmly in place, it moves around. This movement increases electrical resistance and generates heat. Outset Medical advised users to inspect power cords before every use for fraying, cracking, melting, warmth, or loose connections. Users must insert cords fully into inlets and wall sockets, avoid extension cords, and stop treatment immediately if overheating occurs.
What it means
Patients use the Tablo Hemodialysis System in clinics or at home to treat acute or chronic kidney failure. Hemodialysis takes over the job of filtering fluids and toxins from the blood when kidneys can no longer perform that function.
An overheating power cord presents two main dangers. First, touching a hot connector can cause first-degree or second-degree burns. Second, overheating can trip circuit breakers or cut power to the console unexpectedly. A sudden loss of electricity can delay, interrupt, or cancel a dialysis session. As of November 10, 2026, Outset Medical reported no serious injuries or deaths connected to the cord flaw.
What the announcement does not tell us
The alert leaves several details unresolved:
- The announcement does not say how many individual Tablo Consoles, TabloCarts, or power cords are affected by this issue.
- The text does not explain whether Outset Medical plans to replace the power cords or redesign the power inlet to fix the loose connection permanently.
- It does not specify how many total overheating incidents or equipment alarms have occurred so far.
- The document does not state when the FDA will complete its review or decide if a formal product recall is required.
How this was written
This analysis was written with AI from the official announcement below, using only facts that announcement states. The quoted wording is FDA MedWatch Safety Alerts’s own. Always check the original before acting on anything here.
Outset Medical Tablo Console and TabloCart with Prefiltration Drawer power cord connections may overheat, posing fire, burn, and treatment interruption risks.
Opens www.fda.gov
More from FDA MedWatch Safety Alerts
- Early Alert: Heraeus Medical Components Removes Vascular Guidewires20 November 2026
- Breckenridge Pharmaceutical, Inc. Issues Voluntary Nationwide Consumer-Level Recall of One Lot of Propranolol Hydrochloride Extended-Release Capsules, USP, 60 mg, Due to Potential Metal Contamination20 November 2026
- Greenwich Rx Issues Voluntary Nationwide Recall of Compounded Sterile Products Due to Lack of Sterility Assurance20 November 2026
- Medical Device Correction Update Luna® G3 BPAP System (Model LG3800-30VT)19 November 2026
- Early Alert: B. Braun Medical Inc. Updates Use Instructions for Caresite Micro Luer Access Device19 November 2026
This content is for informational purposes only and is not medical advice. It is not intended to diagnose, treat, cure, or prevent any disease. Consult a healthcare professional before starting any supplement, treatment, or program: especially if you are pregnant, nursing, taking medication, or managing a health condition.