FDA Classifies Cardinal Health Instruction Update for Presource Kits as Class I Recall
On November 9, 2026, the U.S. Food and Drug Administration classified a safety instruction update from Cardinal Health regarding certain Presource convenience kits as a Class I recall, the agency's most serious recall category. Convenience kits are pre-packaged sets of medical tools and materials assembled for specific procedures. Cardinal Health issued updated instructions rather than pulling the kits from the market. The affected packs contain BD Alaris Pump Infusion Sets, which deliver fluids and medications to patients. Testing showed that some infusion sets perform worse than previously claimed, especially at flow rates below 1 mL/h or bolus volumes of 5 mL or less. These performance issues can cause incorrect dosing or delayed occlusion alarms, posing risks to vulnerable patients like neonates and critically ill individuals. As of August 19, 2026, BD reported no serious injuries or deaths connected to this issue. The FDA advises healthcare providers to review updated performance data and adjust device usage.
What was announced
The U.S. Food and Drug Administration announced on November 9, 2026, that a safety correction issued by Cardinal Health is now classified as a Class I recall. This classification means using the device without following updated instructions could lead to serious injury or death.
Cardinal Health originally sent letters to affected customers on September 23. The notice affects certain Presource convenience kits, which contain packaged supplies for surgical or medical procedures. These kits include BD Alaris Pump Infusion Sets manufactured by BD and CareFusion 303, Inc. The action does not require facilities to return or remove the kits. Instead, Cardinal Health is instructing users to follow updated risk mitigation steps provided by BD.
What it means
Infusion pumps deliver precise amounts of fluids and medications to patients. Testing by BD revealed that specific infusion sets with restrictive parts, such as filters or check valves, do not deliver fluid as accurately as previously claimed. The problems occur most frequently at low flow rates under 1 mL/h and small bolus volumes at or below 5 mL. A bolus is a single, concentrated dose of medication given over a short period.
These inaccuracies may lead to under-infusion, over-infusion, or delayed warnings when tubing becomes blocked. High-risk patients receiving small medication doses, such as infants or critically ill individuals, face the highest risk of harm.
Healthcare providers using these kits should review the performance disclosures sent by BD. Recommendations include hanging solution containers no higher than 21 inches above the pump module, keeping clamps and vents open, and turning fluid flow off before clearing blockages. If the pump is not suitable for a patient's clinical needs, clinicians should switch to an alternative device, such as a syringe pump.
What the announcement does not tell us
The announcement does not state how many Presource kits or infusion sets are affected by this recall. It also does not specify which hospital departments or geographic regions received the impacted products.
While BD reported zero serious injuries or deaths associated with the issue as of August 19, 2026, the document does not clarify whether any adverse events were reported after that date. Finally, the notice does not mention a timeline for when BD or Cardinal Health might release redesigned infusion sets to fix the underlying performance issues permanently.
How this was written
This analysis was written with AI from the official announcement below, using only facts that announcement states. The quoted wording is FDA MedWatch Safety Alerts’s own. Always check the original before acting on anything here.
Cardinal Health Presource Kits may contain BD Alaris Pump Infusion Sets with Updated Use Instructions.
Opens www.fda.gov
More from FDA MedWatch Safety Alerts
- Game Changer Products LLC Issues Voluntary Nationwide Recall of 10 Day Forecast Capsules Due to the Presence of Undeclared Sildenafil and Acetaminophen12 November 2026
- Early Alert: Intersurgical Inc. Removes 22mm Smoothbore breathing tubes and systems10 November 2026
- Early Alert: Arrow International Removes Products Containing Spring Wire Guide Components10 November 2026
- Heart Pump Controller Recall: Abiomed Removes Automated Impella Controllers (AIC)9 November 2026
- Eugia US LLC (f/k/a Auro Medics Pharma LLC) Issues Voluntary Nationwide Recall of Vancomycin Hydrochloride for Injection USP 1.25g/Vial-30mL (Single Dose Vial) Due to Presence of Particulate Matter9 November 2026
This content is for informational purposes only and is not medical advice. It is not intended to diagnose, treat, cure, or prevent any disease. Consult a healthcare professional before starting any supplement, treatment, or program: especially if you are pregnant, nursing, taking medication, or managing a health condition.